You list several Q-Plex™ assay kits that seem to require a special reading device and software. Can I use an ELISA plate reader for these kits?
A:
The Q-Plex™ kits have a luminescent output. Standard absorbance-based plate readers, typically used for ELISA kits, cannot measure the output of these kits. Each well of these assays contains a matrix of antibodies designed to capture multiple signals from each well. Therefore, even a luminescent plate reader may not be able to capture multiple signals. The Q-View™ Imager LS has been optimized to capture this data. The onboard Q-View™ Software is optimized with this viewer. If you have further questions, contact our Technical Support scientists.